Genauigkeit der fluoroskopischen navigation von pedikelschrauben: CT-basierte evaluierung der schraubenlage

Translated title of the contribution: Accuracy of fluoroscopic navigation of pedicle screws. Assessment of screw placement by computed tomography

E. Fritsch*, J. Duchow, R. Seil, I. Grunwald, W. Reith

*Corresponding author for this work

Research output: Contribution to journalArticleResearchpeer-review

31 Citations (Scopus)

Abstract

While the advantages of C-arm navigation in computer-assisted spine surgery are obvious, the accuracy of pedicle screw placement with virtual fluoroscopy still needs to be verified. The C-arm-based ION system (Medtronic Sofamor Danek) was used to navigate pedicle screw insertion in patients undergoing spinal surgery for various conditions. In a prospective study, a total of 160 screws were inserted in the first 30 consecutive patients since introduction of the system at our institution: 54 at the thoracic spine (highest level: TH4) and 106 at the lumbar spine. Computed tomography (CT) scans were performed postoperatively by two independent radiologists to control the accuracy of screw placement at the level of the pedicles after reconstruction of axial images according to Laine et al. The comparison of the calculated accuracy rate of pedicle screw placement using virtual fluoroscopy with reported results achieved with CT-based navigation shows similar results for virtual fluoroscopy and a remarkable increase of accuracy in comparison to reports on conventional pedicle screw placement.

Translated title of the contributionAccuracy of fluoroscopic navigation of pedicle screws. Assessment of screw placement by computed tomography
Original languageGerman
Pages (from-to)385-391
Number of pages7
JournalDer Orthopade
Volume31
Issue number4
DOIs
Publication statusPublished - Apr 2002
Externally publishedYes

Keywords

  • Accuracy
  • Computer-assisted surgery
  • Fluoroscopic navigation
  • Pedicle screws
  • Spine

Fingerprint

Dive into the research topics of 'Accuracy of fluoroscopic navigation of pedicle screws. Assessment of screw placement by computed tomography: CT-basierte evaluierung der schraubenlage'. Together they form a unique fingerprint.

Cite this